Palm A/B Ultrasound

Micro-Instruments
Disposable Products

Excimer Laser

Green Laser

YAG Laser

Microscope
Phaco

Automated Perimeter

Corneal Topographer

Phaco System Specifications

Mode

Specification

Function

Specification

IRR

Irrigation only, gravity feed

Panel

Mode enabled at fully preset values

I/A

Adjustable aspiration rate
(Rate): 0 to 40 cc/min
Adjustable vacuum limit (I/A VAC): 0 to 500 mmHG
 

Surgeon

Footpedal control of aspiration rate in I/A, VIT
Footpedal control of U/S power in U/S modes

U/S STD

Adjustable U/S power
(U/S Rate): 0 to 100%, 40 kHz
Adjustable aspiration rate
(Rate): 0 to 40 cc/min
Adjustable vacuum limit
(U/S VAC): 0 to 200 mmHg
 

Cautery  

0 to 100%,
10 W into 600 ohms, 50 W max
 

U/S Pulse

As above except U/S power is
pulsed at 10 pulses/sec
 

Reflux

Rate of 25 cc/min

VIT

Adjustable cut rate
(VIT): 60 to 600 cuts/min, single cut
Adjustable aspiration rate
(Rate): 0 to 40 cc/min
Adjustable vacuum limit
(U/S VAC): 0 to 200 mmHg
 

Audio

Audible tone, frequency increases with vacuum

Electrical

100 to 240 VAC, 50 to 60 Hz  

Display

Specification

 

Console:  

Timer

Accumulated U/S time w/rese

 

Depth: 32 cm (13 in)
Height: 11 cm (4.4 in.)
 

VAC

Preset vacuum with footpedal at rest
Actual vacuum with footpedal depressed

 

Width: 40 cm (16 in)
Weight: 8.4 kg (18.5 lbs)
 

U/S Power

Preset U/S power with footpedal at rest
Actual U/S power with footpedal




TEL:
86-24-23852858
86-24-8350 3116
FAX: 86-24-8350 3117
Postal Address: Room 4812, 9#, 11 Wei Road,
Shenyang,China 110000
Sales Manager: sales@ebemp.com

Copyright (c) 1999-2005

1999-2005EBE Sci&Tra co.,Ltd.